Malrotation Due to a Design Element of a New Antegrade Femoral Nail
M Suthersan, I Harris, A Suzuki
Citation
M Suthersan, I Harris, A Suzuki. Malrotation Due to a Design Element of a New Antegrade Femoral Nail. The Internet Journal of Orthopedic Surgery. 2012 Volume 19 Number 3.
Abstract
The authors would like to acknowledge the Synthes company for providing the Expert A2FN nail and set as well as the Synbone femurs to complete this experiment.
Introduction
Intramedullary (IM) nailing has now become the preferred method of treatment for the majority of femoral diaphyseal fractures. Interlocking provides rotational stability and maintains length, thus allowing early movement, and avoiding mal-union of the fracture1,2. However, the closed approach makes anatomical reduction under direct vision impossible, resulting in less rotational accuracy when compared to the traditional method of plate fixation1,2,3. Mal-rotation is a well-established complication of IM nailing of the femur. Several studies have shown the incidence of rotational deformities after femoral nailing of 10 degrees or greater in up to 20% of cases4,5,6. In recent years there have been several developments in the design of antegrade femoral nails. We seek to identify a new design element, spiral grooves along the body of a new antegrade femoral nail (previous grooves were parallel with the nail), as a cause of mal-rotation.
This feature was brought to our attention after a case of malrotation following insertion of a Synthes Antegrade Femoral Nail (A2FN) in a 32 year old patient with a comminuted fracture of the proximal femoral shaft. After completing the procedure, we identified a significant rotational deformity of approximately 45 degrees. Under the same anesthetic, we removed the distal screws and manually corrected the deformity.
This case has led us to examine the design of the A2FN- in particular, the spiral grooves which line nearly the entire length of the nail. It is our hypothesis that the spiral grooves have the ability to grip the femur, particularly at the isthmus, and that, as the nail is inserted, it will grip and rotate the distal fragment. We feel that this is more likely to occur with high femoral shaft fractures (above the isthmus) particularly sub-trochanteric fractures, for which this nail is often used
Materials and Methods
Five synthetic femora (SYNBONE Solid Foam models) were cut transversely at an identical sub-trochanteric point to simulate a sub-trochanteric fracture. A large bolt was then inserted distally into the medullary canal with a large washer, and the bolt was fixed with a clamp, so that the femur was oriented vertically. The femur was free to rotate on the washer and the bolt acted as the centre of rotation. A 3.0mm guide wire was drilled into the bone at an identical distal point, such that the guide wire was pointing anteriorly. The femoral canal was then reamed to 10mm from a trochanteric entry point. Due to the artificial nature of the model, over-reaming did not allow any grip between the nail and the canal, which is usually encountered in vivo.
A 340mm by 10mm A2FN was prepared with another 3.0mm guide wire fixed into the proximal locking hole. The impaction device was then attached to the nail. An initial photograph was taken of the nail as it sat in the femoral canal prior to insertion (Figure 1). This photograph was oriented so that the angle between the two guide wires was 90 degrees. Once the nail had been inserted to the end of the spiral grooves, a final photograph was taken from a similar position (Figure 2) and the rotation between the two wires was measured. The nail was inserted without rotation during insertion.
Results
Internal rotation occurred in each case, due to the gripping of the grooves of the nail in the isthmus of the bone. The degree of internal rotation was 51, 56, 62, 65 and 71 degrees in each case. The average internal rotation was 61 degrees (range: 51 to 71 degrees).
Figure 1
Figure 2
Discussion
Our initial concern was that the new spiral grooves would cause rotation of the distal fragment during insertion, if the nail was inserted without sufficient rotation. Our experiment confirms that if the A2FN is able to grip the femur it can cause a rotational deformity. We believe that this is less likely if the grooves along the nail are straight, or if the grooves did not exist at all (as with other nails). The rotation can be minimised by rotating the nail during insertion, corresponding to the rotation in the spiral grooves. However, ‘straight’ insertion of this nail has the potential to cause mal-rotation. Depending on the grip of the nail in the bone, correction of the rotation around the nail is possible, once any mal-rotation has been identified.
The degree of potential rotation is likely to be higher than our estimate, as we did not completely insert the nail during our experiment. Surgeons should be aware of the large degree of potential rotation associated with straight insertion of this particular nail and, as with any femoral nailing, care should be taken to assess rotation at the end of the procedure.