Forensic Detection Of Fire Accelerants Using A New Solid Phase Microextraction (SPME) Fiber
U Ahmad, G Selvaraju
Keywords
accelerants, arson, fire debris, hs-spme, lab-made c8-coated fiber, n-alkane hydrocarbons
Citation
U Ahmad, G Selvaraju. Forensic Detection Of Fire Accelerants Using A New Solid Phase Microextraction (SPME) Fiber. The Internet Journal of Forensic Science. 2006 Volume 2 Number 1.
Abstract
An important aspect of an investigation of a suspected arson case involves the chemical analysis of the debris remaining after the fire. Currently, accelerant extraction and analytical techniques have been refined to improve sample turnover and to reduce the number of inconclusive findings. For this purpose, solid-phase micro extraction (SPME) have been introduced. SPME relies upon the concentration of headspace vapors onto an adsorbent medium. A new lab-made fiber prepared by sol-gel method, containing 1:1 molar ratio of octyltriethoxysilane (C8-TEOS): methyltrimethoxysilane (MTMOS) was employed in this technique. The fiber was evaluated for the analysis of n-alkane standard hydrocarbon compounds and common petroleum based accelerants. Compared with commercial PDMS/DVB fibre, the new lab-made fibre exhibited higher extraction capability for n-alkane hydrocarbon compounds and accelerants, higher thermal stability (up to 300 °C) and longer lifetime (~ 200 times usage). Electron microscopy experiments revealed that the surface of the fiber coating was well-distributed and a porous structure was suggested for the sol-gel derived C8 coating with an approximate thickness of (3-4) µm. The developed HS-SPME method using C8-coated fiber showed satisfactory reproducibility (RSD < 6 %), detection limits for accelerants (0.7-1.0 µL) and linearity ((r > 0.9869) under the optimum experimental conditions.
1. Introduction
The forensic discipline of ignitable liquid and fire debris analysis is rapidly changing. Modifications to traditional ignitable liquid extraction methods and research into new applications of existing extraction techniques continue to improve the quality of arson analysis. The earliest methods of identification relied upon simple identification of the headspace odor-often referred to as a “nasal appraisal”. With improvements in analytical instrumentation, vapor samples were taken and subjected to instrumental analysis, usually gas chromatography. A significant improvement came with the application of steam distillation to fire debris analysis and this was followed by passive and dynamic headspace method [1, 2].
In addition to optimizing existing extraction techniques, the field of forensic science has also introduced a new extraction technique for application to arson analysis: solid-phase microextraction (SPME). The major advantage of this technique is that it uses no solvents and can be used for either direct sampling or sample clean-up. It is fairly economical and is a relatively simple and sensitive technique. The extraction is based on the enrichment of components on a polymer or adsorbent coated fused silica fiber [3,4,5,6].
SPME is predominantly performed on commercial SPME fibers; however the extent of selectivity obtainable using conventional fibers is limited. Generally accepted drawbacks of conventional adsorbents are a relatively low thermal stability (200-270) ° C which leads to incomplete sample desorption and sample carry-over problem, short lifetime (40-100) times, poor solvent stability and expensive [7,8,9,10,11,12,13,14,15,16,17,18,19,20,21,22,23,24,25].
Recently, many novel coatings have been developed using different techniques and technology for use in SPME. Compared with commercially available SPME adsorbents, the new materials exhibited higher thermal stability (350 ° C), solvent stability, extraction capability and longer lifetime [7,8,9,10,11,12,13,14,15,16,17,18,19,20,21,22,23,24,25]. However, up to now, none of the novel fibers have been evaluated for the determination of accelerants in arson analysis. This paper presents a recent development in the forensic aspects of fire investigation. As a preliminary study, a new SPME adsorbent comprising of sol-gel derived C8-coating was developed and evaluated for the determination of accelerants in arson samples, with the aim of improving the quality of ignitable liquid residue analysis.
2. Experimental
2.1. Chemicals and Materials
Individual standards of
2.2. Apparatus
Two glass apparatus (400 cm 3 and 125 cm 3 ) for sample preparation step of HS-SPME was specially designed [26]. A Supelco SPME holder, commercially available PDMS/DVB fiber (Bellefonte, Pennsylvania, U.S.A.) and a used SPME fibre with a burnt off tip, coated with sol-gel derived C8-coated fiber containing (1:1, C8-TEOS:MTMOS) [27] were employed for the extraction of
2.3. Instrumentation
Gas chromatography analyses were conducted using a Hewlett-Packard 6890 GC (Wilmington, Delaware, U.S.A.). The HP 6890 gas chromatograph was equipped with FID and a HP ChemStation for data processing. An Ultra-1 capillary column (Agilent) of dimensions 25 m x 0.20 mm x 0.11 µm film thickness was used. Helium was used as the carrier gas at a flow rate of 1.2 mL/min. The injection port temperature was set at 250 °C and FID temperature at 310 °C. SPME injections were performed using a split mode injection (5:1).
Studies of fiber coating structure and thickness were made by means of Philips Scanning Electron Microscope model XL 30 SEM (Philips Electronic Instruments Company, Mahwah, New Jersey) equipped with a ThermoNoran energy dispersive X-ray detection system (EDX).
2.4. Procedures for Headspace SPME
30 µL from the prepared standard solution of
2.5. Preparation of Fire Debris sample
A sample of carpet (20 cm x 13 cm) placed on a sheet of aluminium foil was ignited with a fire starter and left to burn until about one-third remained on the aluminium foil. Fire was extinguished by cutting off the oxygen supply. The partially burnt carpet was then exposed to the surrounding air for 30 minutes to let it cool down.
3. Results and Discussion
3.1. Characterization of the C-coated Fiber
3.1.1. Selectivity for n-alkane Standard Hydrocarbons
In order to examine selectivity of the lab-made SPME fiber towards hydrocarbon compounds, a mixture of
3.1.2. Extraction Capability for n-alkane Standard Hydrocarbons
The extraction capability of the C8-coated fiber for hydrocarbons was determined by comparing it with the extraction capability of commercially available PDMS/DVB fiber. PDMS/DVB fiber was selected for comparison because previous work done in this lab [26] proved that the fiber has the highest sensitivity towards hydrocarbon compounds. As can be seen from Figure 2, the C8-coated fiber exhibited a slightly higher extraction capability for all the hydrocarbon compounds by contrast with conventional PDMS/DVB fiber. A higher extraction capability yielded by C8-coated fiber could be due to the existence of higher surface area for the C8-coated fibers [13].
3.1.3. Lifetime of the C-coated Fiber
The long lastingness of the C8-coated fiber was determined by 200 continued operations carried out with the same fiber and oven temperature program. The C8-coated fiber have been used for hydrocarbon extractions (C9, C13, and C15) and subjected to GC for more than 200 times. There was no significant differences of hydrocarbon peaks obtained in each operation. All the hydrocarbon compounds gave a low relative standard deviation (RSD) value ranging from 3.8 %-5.4 % which shows an acceptable reproducibility. This proves that the coated surface of the fiber was not partially depleted during the continued operation. It was still stable and reusable. Such a long service life are possibly due to the strong chemical bonding between the sol-gel generated C8-coated composite coating and the silica surface [7,12,13, 15].
3.1.4. Thermal Stability of the C-coated Fiber
The effect of conditioning temperature on the stability of the C8-coated fiber was determined by conditioning the fiber at high temperatures (270 and 300) ° C for 1 hour prior to extraction. High temperature conditioning lead to consistent improvement in peak area repeatability for SPME-GC analysis. The RSD value of < 4.6 % can be routinely obtained for hydrocarbons on C8-coated fibers conditioned at 270 ° C and 300 ° C. The C8-coated fiber can be routinely used at 300 ° C without any sign of bleeding, whereas for commercial PDMS/DVB fiber, the highest temperature the coating layer can endure is less than 280 ° C, and thus the range of analyte molecular weights that can be handled by SPME-GC is limited. Enhanced thermal stability of C8-coated fiber might be due to the strong chemical bonding between the sol-gel generated composite coating and the silica surface [8, 11, 16, 17,18,19, 24, 25].
3.1.5. Scanning Electron Microscopy (SEM) Analysis
The morphology of the lab-made fiber was investigated using the scanning electron microscope. As can be seen from Figure 3, it is obvious that the surface of the fiber coating was well-distributed. A homogeneous porous structure was suggested for sol-gel C8-coating (Figure 4). Chong
The average thickness of the C8 coating was estimated as being ~ 3-4 µm, which is considerably thinner than most conventional SPME fibers (65 µm for PDMS/DVB). The reproducibility of the coating thickness obtained in this study is rather promising. The volume of extracting phase on the lab-made C8-coated fiber is ~ 0.011-0.014 mm 3 whereas a 65 µm PDMS/DVB commercial fiber has an extracting phase volume of 0.357 mm 3 . It is apparent that the volume of extracting phase on the lab-made C8-coated fiber is considerably less than that of commercial PDMS/DVB fiber. Therefore the C8 coating has a larger affinity towards the analytes than regular PDMS/DVB.
3.2. Method Validation
3.2.1. Accuracy of HS-SPME Method using C-coated Fiber
The accuracy of the developed method for three replicate analyses was determined by calculating the relative standard deviations (RSD) of the peak areas for all standard hydrocarbons (C8, C10, C12, C14 and C16) in burnt carpet sample spiked at the same concentration level.
All the hydrocarbon compounds gave a low relative standard deviation (RSD) value ranging from 3.9 %-5.5 % which shows an acceptable reproducibility. These results indicated that quantitative analysis of hydrocarbons was possible using the external standard method and the applicability of lab-made C8-coated fiber for routine analysis.
3.2.2. Detection Limits of Accelerants
Detection limits of accelerants (lowest detectable amount determined based on three times the baseline noise) were evaluated in order to access the performance of the proposed method. The lab-made C8-coated fiber was employed for extractions of gasoline, kerosene and diesel in spiked burnt carpet sample under optimum HS-SPME conditions.
For all the common accelerants, the LODs were between 0.7 to 1.0 µL. These values are indicative of good performance of the lab-made C8-coated fiber. The detection limits of common accelerants obtained is this study using C8-fiber is comparable to those reported by Yong [26] using commercial PDMS/DVB fiber.
3.2.3. Linearity
Calibration curves consisting of five different concentration of
HS-SPME procedure with C8-coated fiber yielded an excellent linear relationship between the peak area counts and analyte concentration in the range of 2.00-10.00 % v/v. Good correlations with R 2 value between 0.9869-0.9958 (R 2 ~ 1) were obtained. It is apparent that there were almost no factors in the HS-SPME extractions of
3.3. Analysis of Simulated Arson Samples using C-coated Fiber
3.3.1. Selectivity for Petroleum Based Accelerants
In order to examine selectivity of the C8-coated fiber towards petroleum based accelerants, burnt carpet sample were individually spiked with known amount of gasoline, kerosene and diesel. The samples were subjected to headspace SPME using C8-coated fiber and the GC profiles were compared with that from direct injection. As seen in Figure 6 (b), all the hydrocarbon components in gasoline spiked burnt carpet sample were recovered by using C8-coated fiber and the chromatogram was comparable with the profile of gasoline from direct injection [Figure 6 (a)]. The hydrocarbon components in kerosene (Figure 7) and diesel (Figure 8) spiked burnt carpet sample were also effectively extracted using the C8-coated fiber and similar comparisons were obtained with direct injection of the respective accelerants. This indicated that the C8-coated fiber favored the extraction of hydrocarbons, thus providing a good selectivity towards accelerants. The GC profiles of gasoline, kerosene and diesel spiked samples obtained in this study were in good agreement with those obtained by Borusiewicz [1] using Tenax as adsorbent and Yong [26] using commercially available fibers. Figure 6: GC profiles of direct injection of gasoline
Figure 6
Figure 7
3.3.2 Extraction Capability for Petroleum Based Accelerants
The extraction capability of the sol-gel derived C8-coated fiber was determined by comparing it with the commercially available PDMS/DVB fiber. Both C8-coated fiber and PDMS/DVB fiber were capable of extracting early, middle and late eluting hydrocarbon compounds sufficiently. However, the C8-coated fiber showed a slightly higher extraction capability by contrast with conventional PDMS/DVB fiber for all the accelerants as shown in Figure 9. This result was comparable to that reported by Gbatu [13]. A higher extraction capability could be due to the existence of higher surface area for the C8-coated fibers [9, 13, 14, 20,21,22,23].
3. Conclusions
A new SPME fiber (lab-made C8-coated fiber) has been successfully developed and evaluated for the determination of common petroleum based accelerants in arson samples. The C8-coated fiber has a good selectivity towards hydrocarbon compounds and accelerants as it was capable of effectively extracting all the tested n-alkane hydrocarbon compounds and hydrocarbon components in gasoline, kerosene and diesel spiked burnt carpet samples. The lab-made fibre gave superior characteristics when compared to commercially available PDMS/DVB fibre. Compared with commercial PDMS/DVB fibre, the new lab-made fibre exhibited higher extraction capability for n-alkane hydrocarbon compounds and accelerants, higher thermal stability (up to 300 ° C) and longer lifetime (~ 200 times usage). Electron microscopy experiments revealed that the surface of the fibre coating was well-distributed but not uniformly coated and a porous structure was suggested for the sol-gel derived C8 coating with an approximate thickness of (3-4) µm. The developed HS-SPME method using C8-coated fiber exhibited a high degree of accuracy under the optimum experimental conditions. The limits of detection for the common accelerants were in the range of 0.7 to 1.0 µL. The linearity of the calibration graphs for all the tested
Acknowledgements
The authors would like to express their deepest gratitude to the Ministry of Science, Technology and Innovation, Malaysia (MOSTI) for the financial support under IRPA RM8 mechanism vote No. 74090.